20
Post-use Maintenance
6-5
Maintenance of the Water Line
for the Motor and the Motor Cord
Cleaning and Disinfection should be done in the cleaning tank.
CAUTION
Be sure to perform Cleaning and Disinfection the "6-3 Maintenance after Close (Cleaning and Disinfection
of the outside of the Motor and the Motor Cord)" before "6-5 Maintenance of the Water Line for the Motor
and the Motor Cord".
6-5-1 Preparation of the Water Line before Disinfection
Disinfect with Alpro Bilpron (ALPRO) (undiluted) is recommended as maintenance once or twice week.
Put 200 mL of disinfectant solution in the Water Bottle. Tightly close the Water Bottle lid and place
the Water Bottle in the Control Unit.
* For details, confirm the operation manual isseued by manufacturer.
Disinfectant Solution: Alpro
Bilpron (ALPRO) (undiluted)
1
Disinfectant Solution
200
mL
· When there is not enough disinfectant solution,
turn off the Power Switch of the Control Unit,
take out the Water Bottle, and add disinfectant
solution.