Post-use Maintenance
Apply autoclave sterilization for the following products. After each patient treatment, sterilize the products
as follow.
Autoclavable equipment: Motor, Motor Cap, Protection Plug,
< Preparation before sterilization >
Tighten the Motor Cap to the connection
part of the Motor Cord and place the
Protection Plug in the insert part of the
Motor.
1
Prote ction Plug Motor Cap
Tighten
Insert the Motor with the Protection Plug
and the Motor Cap to the sterilization pouch
that conform to ISO 11607-1 and seal the
pouch.
2
< Sterilization >
Autoclave the Motor. Conduct autoclave sterilization as follow.
Type Gravity Displacement Pre-Vacuum (Dynamic Air Removal)
Temperature 121
°C (
0 / +4
°C)
132
°C
(0 / +4
°C)
134
°C
(0 / +4
°C)
Full Cycle Time 20 min. or longer 15 min or longer 3-18 min
Drying Time 30 min or longer 30 min or longer 30 min or longer
Store the product in a place where it is kept clean and keep it in a sterilization pouch until it is used next.