Post-use Maintenance
6-4 Sterilization
CAUTION
Follow local rules, regulations, and guidelines regarding the reprocessing of devices.
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible
discoloration and damage to the product from chemical residue on other instruments.
Be sure to use sterilizers that can perform sterilization up to 135°C. In some sterilizers, the chamber
temperature may exceed 135°C. Do not use these sterilizers as failure of the product could occur. Contact
the sterilizer manufacturer for detailed information about cycle temperatures.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The
air should be free from dust, salt and sulphur.
Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile
condition.
Autoclave sterilization is recommended for the product. The validity of other sterilization methods (such as
plasma sterilization or EOG sterilization) is not confirmed.
Immediately after sterilization is complete (within 1 hour), remove the Motor from the sterilozer. Failure to
do so may cause corrotion.
NOTICE
NSK recommends Class B sterilizer as stated in EN 13060.
For details, confirm to the manufacturer's of the sterilizer operation manuals.