22
6-5-2
Disinfection of the Water Line of
the Motor and the Motor Cord
Push the MOTOR Key.
Turn ON the SPRAY Key to irrigation.
1
Hold the Motor and the Motor Cord in the
cleaning tank.
2
Press the ON/OFF Key for 2 seconds and
activate for
10 seconds or more to confirm
that the disinfectant solution comes out
from the tip of the Motor.
3
Place the Motor and the Motor Cord on the
Holder of the Control Unit state for a night (at
least 12 hours).
4
Post-use Maintenance
CAUTION
•
Disinfectant solution remains in the Water Line by letting the solution out from the tip of the Motor.
•
It takes 12 hours to complete the disinfection after disinfectant solution remain on the Water Line of the
Motor and the Motor Cord.