CAUTION
Chuck cleaning
Clean the Push Button chuck once a week.
1) Mount the Tip Nozzle into the spray can port.
2) Lubricate the chuck directly through the bur insertion hole (Fig. 10).
3) Lubricate handpiece by using NSK PANA SPRAY Plus/PANA SPRAY (Fig. 9) or NSK automatic handpiece
cleaning and lubrication system.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system, refer to the system instructions.
Ř If
the chuck is not regularly cleaned the chuck grip may be weakened and the bur may be accidentally released while in operation.