EasyManuals Logo

NSK Ti-Max Z Series User Manual

NSK Ti-Max Z Series
146 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
15
EN DE FR ES IT PT
CAUTION
8-6 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
ŘDo not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
ŘKeep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
ŘImmediately after use, the product should be cleaned, lubricated and sterilized. If blood remains on the external or internal surfaces it can
become clotted and cause rust.
ŘDo not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
ŘIf the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the dry cycle.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NSK Ti-Max Z Series and is the answer not in the manual?

NSK Ti-Max Z Series Specifications

General IconGeneral
BrandNSK
ModelTi-Max Z Series
CategoryDental equipment
LanguageEnglish

Related product manuals