20
13
Specifications
<Control Unit>
<Foot Control>
<Handpiece>
Model VarioSurg 3
Rated voltage AC120V 50/60Hz
AC230V 50/60Hz
Drive frequency 28 - 32kHz
Maximum output 25W
Power input 54VA
Dimension W265 x D220 x H103mm
Weight 3kg
Model FC-78
Cord Length 2m
Dimension W268 x D230 x H103mm
Weight 1.4kg
Model VS3-LED-HPSC
Cord Length 2m
Type of oscillation Piezo type
Optic White LED
Consumption current(LED) 0.17A (3.5V)
Dimension ø20×129mm (except cord)
Weight 70g (without cord)
Temperature Humidity Atmospheric pressure
Use environment 0 - 40°C 30 - 75% -
Transportation and Storage
environment
-10 - 50°C 10 - 85% 500 - 1060hPa
* No liquid freezing allowed
* When operated outside the above range, there is a risk of failure.
14
Classification of Equipment
ŘType of protection against electric shock
- Class I equipment
ŘDegree of protection against electric shock:
- Type BF applied part (Applied part: Tip, Handpiece)
ŘMethod of sterilization or disinfection recommended by the manufacturer:
- See”9-6 Sterilization”
ŘDegree of protection against ingress of water as detailed in the current edition of IEC 60529:
- Foot Control: IPX8 (Protected against the effects of continuous immersion in water)
ŘDegree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide:
- Equipment NOT suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide. (Without Foot Control)
ŘFoot Control: Category AP Equipment
ŘMode of operation:
- Continuous operation
15
Operation Principle
A sinusoidal electrical signal, at ultrasonic frequency (f>20kHz), is delivered by the generator. This signal is applied to the
‘piezoelectric ceramic’ located inside the transducer. Piezoelectric ceramic converts this signal into mechanical vibrations.
These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations are propagated
towards the distal end of the transducer. The “TIP” insert, which is attached at the distal end of the transducer, vibrates at
ultrasonic frequencies and makes it possible to achieve the aimed purpose.