22
18
Option Parts List
19
Spare Parts List
Model Order Code Remarks
SG Link Set Y1002729
Model Order Code Remarks
LED handpiece with Cord E1133001
Irrigation Tube Y900113 5 pcs
Tube Holder Y900767 7 pcs
Tips Holder 20001327
Tip Wrench 10000977
E Tip Wrench Z217399 For V10-S
20
Disposal of the product
21
EMC Information (Electromagnetic Compatibility Information)
In order to avoid the health risks of operators handling the disposal of medical equipment, as well as the risks of
environmental contamination caused thereof, a surgeon or a dentist is required to confirm the equipment is sterile. Ask
specialist firms who are licensed to dispose of specially controlled industrial wastes, to dispose the product for you.
Guidance and manufacturer's declaration - Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR11/EN55011
RF emissions
CISPR11/EN55011
Harmonic emissions
EN/IEC61000-3-2
Voltage fluctuations/flicker emissions
EN/IEC61000-3-3
Compliance
Group 1
Class B
Not Applicable
Not Applicable
Electromagnetic environment - guidance
The product uses RF energy only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
The product is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration - Electromagnetic Immunity
NOTE: ‘Ut’ is the AC mains voltage prior to application of the test level.
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
EN/IEC61000-4-2
Electrical fast transient/burst
EN/IEC61000-4-4
Surge
EN/IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN/IEC61000-4-11
Power frequency (50/60Hz)
magnetic field
EN/IEC61000-4-8
EN/IEC60601 test level
±(2, 4) 6kV contact
±(2, 4) 8kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(<95% dip in Ut)
for 5 sec
3 A/m
Compliance level
±(2, 4) 6kV contact
±(2, 4) 8kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(<95% dip in Ut)
for 5 sec
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the product requires
continued operation during power mains interruptions, it is
recommended that the product be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
9DULR6XUJ(1LQGG