2
WARNING
ŘDo not handle the power cord with wet hands. Wet hand contact with electricity may result in an electric shock.
ŘKeep away from explosive substances and flammable materials.
ŘIf the product overheats or smells of burning, immediately turn off the power and disconnect the main power plug.
Contact your Authorised NSK Dealer.
ŘTO PREVENT ELECTRIC SHOCK, use a main electrical outlet that is protective earthed.
ŘDo not use on the following patients.
-Those with medical complications or allergies
-Those who have pre existing conditions
(Eg Cardiac, Pulmonary, Renal disturbance or High blood pressure)
-Those who are pregnant or lactating
-Patients with cardiac pacemakers and infants
ŘBe careful not to get water or liquid disinfectant on the control unit. This could cause short circuits and lead to fire and/
or electric shock.
ŘRepeatedly turning the Main Power Switch ON and OFF may blow a fuse.
ŘWhen installing the product, provide space of approximately 10cm around the product for easy access to the inlet and
the Power Cord.
ŘThe Irrigation Tubes, included in the product package, are sterile. When using these products, follow the instructions
below.
- The irrigation tube is a single-use item. Do not reuse nor sterillize as product breakage or infection could occur.
- Check before use that the sterile package is NOT open nor damaged. Do not use products with open or damaged
packages as sterility will be negated and infection could occur.
- Observe the use-by date written on the sterile package label. Do not use expired products as sterility may be
negated.
ŘDo not point the handpiece LED light directly to the eyes of the patient, nor the operator, as it may be harmful to eyes.
CAUTION
ŘTo be confident in using this product, please read this Operation Manual before use. We suggest that to fully
understand the product functions and file for future reference.
ŘWhen operating the product always consider the safety of the patient.
ŘThe end user shall be responsible for any judgment that relates to the application of this product to a patient.
ŘThis product does not consider patient’s age (except infants), gender, weight or nationality.
ŘThis product does not consider operator’s age (mature person), height, weight, gender, or nationality.
ŘUsers are responsible for the operational control, maintenance and continual inspection of this product.
ŘThis device is for indoor use only.
ŘKeep the Control Unit on a level surface.
ŘDo not attempt to disassemble the product or tamper with the mechanism except as recommend by NSK in this
Operation Manual.
ŘDo not allow any impact on to the handpiece. Do not drop the handpiece.
ŘOperators and all others in the area must wear eye protection and a mask when operating this handpiece.
ŘShould the product function abnormally, cease operation immediately and contact your Authorized NSK Dealer.
ŘIf the Irrigation Pump gets wet, wipe it dry. If it remains wet, the roller in the pump may slip and the pump may not
operate properly.
ŘDo not bend or fold the Irrigation Tube while the Irrigation Pump is operating as the Tube may be damaged or may
detach.
ŘIf abnormality is detected in irrigation, it may be due to wear of the Irrigation Tube or leaking of saline from the Tube. In
such cases, replace the Irrigation Tube.
ŘDo not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
ŘThe following products are delivered in a non-sterile condition and must be autoclaved prior to use. (Handpiece, Tip, Tip
Wrench, Tip Holder, Tube Holder)
ŘPerform regular function and maintenance checks.
ŘIf the product is not used for a long period check it is functioning correctly before using on a patient.
ŘTo avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown during surgery.
ŘThis product may be affected by an electric scalpel. Turn OFF the Power Switch when an electric scalpel is used.