NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head. Instruction Manual
Fig. 2-10. LBDS Input socket mounted
2. Secure the Input socket by screwing two fastening screws with Allen key 1.5 mm
(
Fig. 2-10).
3. Close the laser source exit shutter if it is open.
4. Take the Output socket of the LBDS with your right hand while, with your left hand,
press and hold the central backing-up screw of the radiation feedthrough. Insert the
Output socket into the slot of the radiation feedthrough against the stop
(see
Fig. 2-11).
Fig. 2-11. Radiation feedthrough
This completes mounting the LBDS for adjusting the detection channel.