NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head. Instruction Manual
Fig. 3-16. Screws for moving the optical tube
1 – X movement screw; 2 – Y movement screw;
3 – Z movement screw
5. Focus the laser beam on the sample surface by moving the optical tube along Z with
the screw 3 (Fig. 3-16) until diameter of the laser spot image is at its minimum
(Fig. 3-17).
Fig. 3-17. Image of the focused laser spot
6. Move the optical tube in the XY plane to place the laser spot slightly below the center
of the screen (Fig. 3-18).
32