Pulse Mode Bolus Volumes (mL)
Breaths per
Minute
1
2
3
4
5
6
7
8
9
10*
15
8.0
16.0
24.0
32.0
40.0
48.0
56.0
65.0
75.0
80.0
20
8.0
16.0
24.0
32.0
40.0
48.0
56.0
65.0
75.0
80.0
25
6.4
12.8
19.2
25.6
32.0
38.4
44.8
52.0
60.0
64.0
30
5.3
10.7
16.0
21.3
26.7
32.0
37.3
43.3
50.0
53.3
35
4.6
9.1
13.7
18.3
22.9
27.4
32.0
37.1
42.9
45.7
40
4.0
8.0
12.0
16.0
20.0
24.0
28.0
32.5
37.5
40.0
*Pulse Mode setting 10 available with the Liberty2 device.
Pulse Mode No Breath Detection
When the device has not detected a breath for 45 seconds, the screen will turn
amber, sound a 3-beep sequence and the text “No Breath” will flash on the screen. If
the device does not detect a breath after an additional 15 seconds, the device will
beep once and switch to Continuous Mode.
No Breath Detection Recovery
While in this condition, the breath detection icon on the top of the screen will turn
red. The device will check for inspiration for 20 seconds every 15 minutes. If a breath
is detected, the device will return to the last Pulse Mode setting. If no breath is
detected, the device will stay in Continuous Mode and check for inspiration every 15
minutes.
The device will return to the last Pulse Mode setting at next power up or if the power
source is changed.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the O2 Concepts LIBERTY2 and is the answer not in the manual?

O2 Concepts LIBERTY2 Specifications

General IconGeneral
BrandO2 Concepts
ModelLIBERTY2
CategoryMedical Equipment
LanguageEnglish

Related product manuals