38
Section 5: User Alerts and Alarms
The functionality of the alarm system is verified automatically by the unit upon
start up by the unit flashing three (3) visual display colors and audible indicators.
There will be delays that are greater than ten (10) seconds inherent to specific
alarms, and these delays are explained in detail for each alarm in the following
tables. To ensure that audible alarm notifications are heard, the maximum
distance the user can be from the device is dependent on surrounding noise
levels. When two alarm conditions occur at the same time, the higher priority
condition will be displayed.
Indefinite Acknowledge (Alarm Mute)
In an alert or alarm condition, pressing the power button momentarily will put the
device into an indefinite acknowledge state or alarm mute. In this state, the
alarm will be silenced, and the screen will stop flashing. The device will stay in
this state until powered off or until a higher-priority alarm is activated.
DO NOT IGNORE ALERTS OR ALARMS