EasyManuals Logo

O2 Concepts LIBERTY2 User Manual

O2 Concepts LIBERTY2
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
10
91% ± 4% (87 - 95%); Measured purity values are
within ± 2% of actual values; There is no variation in
purity within the operating altitude. This applies to
the full range of environmental operating conditions.
50°F (10°C) to 104°F (40°C)
10% - 95% @ 82.4°F (28°C)
1060hPa to 630hPa
Free of smoke, pollutants, and fumes.
-4°F (-20°C) to 140°F (60°C)
0 -95% non-condensing
24 hours per day when connected to an external
AC or DC power source. This is a continuous
operation device.
DO NOT use cannula tubing longer than
7ft (2.13m) when using Pulse Flow Mode.
DO NOT use cannula tubing longer than 50ft
(15.25m) when using Continuous Flow Mode.
The volume that the device will reach at maximum
settings will be ~ 52.3 dBA. The volume at setting 2
Pulse will be ~ 44.4 dBA.
All alarms triggered by the device will be 85 dBA
or higher at 10cm from the unit.
Maximum outlet pressure is 7.4 psig (51kPa)
Device: 5yrs, Accessories: 1yr

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the O2 Concepts LIBERTY2 and is the answer not in the manual?

O2 Concepts LIBERTY2 Specifications

General IconGeneral
BrandO2 Concepts
ModelLIBERTY2
CategoryMedical Equipment
LanguageEnglish

Related product manuals