i
2000 Series Indicators
Instruction Manual
T24PE Indicator
Manual_Indicator_T24PE_V1

Questions and Answers

  • L
    Leslie MelendezAug 13, 2025
    What does 'CAL E' mean on OHAUS Measuring Instruments?
    • D
      Diane PerezAug 13, 2025
      The 'CAL E' message on OHAUS Measuring Instruments indicates that the calibration value is outside of allowable limits.
  • W
    william87Aug 17, 2025
    Why is the battery icon flashing on my OHAUS Measuring Instruments?
    • U
      uhesterAug 17, 2025
      The battery icon flashes on your OHAUS Measuring Instruments because the battery is empty.
  • W
    wbutlerAug 21, 2025
    What does Error 8.4 mean on OHAUS Measuring Instruments?
    • V
      vjonesAug 21, 2025
      Error 8.4 on OHAUS Measuring Instruments means the weight reading is below the underload limit.
  • R
    Robert HarrisonAug 25, 2025
    What does 'rEF.Err' mean on OHAUS Measuring Instruments?
    • R
      Richard JonesAug 25, 2025
      The 'rEF.Err' message on OHAUS Measuring Instruments indicates that the reference weight is too small. Specifically, the weight on the pan is too small to define a valid reference weight.
  • A
    Amy LarsonAug 28, 2025
    What does Error 8.1 mean on OHAUS Measuring Instruments?
    • R
      Randy FrazierAug 28, 2025
      Error 8.1 on your OHAUS Measuring Instruments indicates that the weight reading exceeds the Power On Zero limit.
  • D
    David CastilloSep 1, 2025
    What does Error 8.2 mean on OHAUS Measuring Instruments?
    • T
      Tony WilliamsSep 1, 2025
      Error 8.2 on your OHAUS Measuring Instruments means that the weight reading is below the Power On Zero limit.
  • J
    Jean ChambersSep 4, 2025
    What does Error 8.5 mean on my OHAUS Measuring Instruments?
    • T
      timothy37Sep 5, 2025
      Error 8.5 on your OHAUS Measuring Instruments means that the device was tared at one unit, but after switching to another unit, the tare value exceeds the maximum.
  • P
    Patrick MercadoSep 9, 2025
    What does 'Lo.rEF' mean on OHAUS Measuring Instruments?
    • K
      kellythompsonSep 9, 2025
      The 'Lo.rEF' message on OHAUS Measuring Instruments is a warning that the Average Piece Weight is too small.

Summary

1. INTRODUCTION

1.2 Safety Precautions

Lists essential safety measures for operating the indicator.

1.3 Overview of Parts and Controls

1.3 T24 PE Indicator Components

Identifies and lists all physical parts of the T24PE indicator.

1.4 Control Functions

1.4 Control Panel Button Operations

Details primary, secondary, and menu functions of the control panel buttons.

2. INSTALLATION

2.2 External Connections

Describes connections for scale base, power, and RS232.

3.3 Calibration Menu

3.3.1 Zero Calibration Procedure

Steps for performing zero calibration on the indicator.

3.3.2 Span Calibration Procedure

Steps for performing span calibration with a calibration mass.

3.3.3 Linearity Calibration Procedure

Steps for performing linearity calibration at three points.

3.3.4 Geographical Adjustment Factor

Explains the GEO factor for gravity compensation and its adjustment.

3.4 Setup Menu

3.4.2 Setting Scale Capacity

Guide to setting the maximum capacity of the scale.

3.6 Mode Menu

3.6.2 Enabling Parts Counting Mode

Enables or disables the parts counting application mode.

3.6.3 Enabling Totalization Mode

Enables or disables the totalization application mode.

3.9 Security Switch

3.9 Security Switch Functionality

Explains the function and setting of the security switch on the PCB.

4. OPERATION

4.1 Turning the Indicator On and Off

Procedure for powering the indicator on and off.

4.2 Zero Operation

Describes methods for zeroing the scale reading.

4.3 Manual Tare Function

How to perform a manual tare operation to exclude container weight.

4.6.2 Parts Counting Mode Operation

Instructions for using the parts counting function.

4.6.3 Totalization Mode Operation

How to measure the cumulative weight of a sequence of items.

6. LEGAL FOR TRADE

6.1 Settings for Legal for Trade

Steps to configure the indicator for legal-for-trade use.

6.2 Verification Requirements

Information on product inspection for trade approval.

7. MAINTENANCE

7.2 Troubleshooting Common Errors

Lists error codes, descriptions, and their causes.

8. TECHNICAL DATA

8.1 Indicator Specifications

Detailed technical specifications of the T24PE indicator.

OHAUS Defender 2000 Series Specifications

General IconGeneral
Readability0.1 g
DisplayLCD with backlight
Linearity±0.2 g
CalibrationExternal calibration
Power SupplyAC Adapter
Weight2.2 lbs (1 kg)
Operating Temperature0 to 40 °C
Units of Measurementg, oz, lb
ConstructionABS plastic housing with stainless steel pan
ApprovalsCE
Repeatability0.1 g
InterfaceRS232