What does 'CAL E' mean on OHAUS Measuring Instruments?
- DDiane PerezAug 13, 2025
The 'CAL E' message on OHAUS Measuring Instruments indicates that the calibration value is outside of allowable limits.
What does 'CAL E' mean on OHAUS Measuring Instruments?
The 'CAL E' message on OHAUS Measuring Instruments indicates that the calibration value is outside of allowable limits.
Why is the battery icon flashing on my OHAUS Measuring Instruments?
The battery icon flashes on your OHAUS Measuring Instruments because the battery is empty.
What does Error 8.4 mean on OHAUS Measuring Instruments?
Error 8.4 on OHAUS Measuring Instruments means the weight reading is below the underload limit.
What does 'rEF.Err' mean on OHAUS Measuring Instruments?
The 'rEF.Err' message on OHAUS Measuring Instruments indicates that the reference weight is too small. Specifically, the weight on the pan is too small to define a valid reference weight.
What does Error 8.1 mean on OHAUS Measuring Instruments?
Error 8.1 on your OHAUS Measuring Instruments indicates that the weight reading exceeds the Power On Zero limit.
What does Error 8.2 mean on OHAUS Measuring Instruments?
Error 8.2 on your OHAUS Measuring Instruments means that the weight reading is below the Power On Zero limit.
What does Error 8.5 mean on my OHAUS Measuring Instruments?
Error 8.5 on your OHAUS Measuring Instruments means that the device was tared at one unit, but after switching to another unit, the tare value exceeds the maximum.
What does 'Lo.rEF' mean on OHAUS Measuring Instruments?
The 'Lo.rEF' message on OHAUS Measuring Instruments is a warning that the Average Piece Weight is too small.
Readability | 0.1 g |
---|---|
Display | LCD with backlight |
Linearity | ±0.2 g |
Calibration | External calibration |
Power Supply | AC Adapter |
Weight | 2.2 lbs (1 kg) |
Operating Temperature | 0 to 40 °C |
Units of Measurement | g, oz, lb |
Construction | ABS plastic housing with stainless steel pan |
Approvals | CE |
Repeatability | 0.1 g |
Interface | RS232 |
Lists essential safety measures for operating the indicator.
Identifies and lists all physical parts of the T24PE indicator.
Details primary, secondary, and menu functions of the control panel buttons.
Describes connections for scale base, power, and RS232.
Steps for performing zero calibration on the indicator.
Steps for performing span calibration with a calibration mass.
Steps for performing linearity calibration at three points.
Explains the GEO factor for gravity compensation and its adjustment.
Guide to setting the maximum capacity of the scale.
Enables or disables the parts counting application mode.
Enables or disables the totalization application mode.
Explains the function and setting of the security switch on the PCB.
Procedure for powering the indicator on and off.
Describes methods for zeroing the scale reading.
How to perform a manual tare operation to exclude container weight.
Instructions for using the parts counting function.
How to measure the cumulative weight of a sequence of items.
Steps to configure the indicator for legal-for-trade use.
Information on product inspection for trade approval.
Lists error codes, descriptions, and their causes.
Detailed technical specifications of the T24PE indicator.