Table 9-5 SPECIFICATIONS (continued)
ExCal Model PR2201N/E PR4201N/E PR5201N/E PR6201N/E
Capacity (g) 2200 4200 5200 6200
Readability d (g) 0.1 0.1 0.1 0.1 or1
Verification interval e(g) 0.1 0.1 0.1 1
Class II II II III
Repeatability (sd.), 5% of
Full Load (g)
0.08 0.08 0.08 0.08
Repeatability (sd.), 5% of Full
Load to Full Range (g)
0.1 0.1 0.1 0.1
Linearity Deviation, Typical
(g)
0.06 0.06 0.06 0.06
Linearity Deviation (g)
0.2 0.2 0.2 0.2
Eccentric Load Not exceeding the maximum permissible error for the one-third of the full capacity of the balance
Stabilization Time Typical (s) 1 1 1 1
Sensitivity Temperature Drift
(
PPM/K
)
±10 ±10 ±3 ±10
Typical Minimum Weight
USP (USP K=2,U=0.10%)
160g 160g 160g 160g
Optimized Min-Weight
(USP, u=0.10%, k=2) 82g 82g 82g 82g
Units Gram, Kilogram, Carat, Pound, Ounce, Ounce Troy, Grain
Applications Basic Weighing; Parts counting; Percent weighing
Platform Size (diameter) 7.1 in / 18 cm
Span Calibration Points (g) 1000, 2000 2000, 4000 3000, 5000 5000, 6000
Linearity Calibration Points
(
g
)
0, 1000, 2000 0, 2000, 4000 0, 2500, 5000 0, 3000, 6000
Tare Range To capacity by subtraction
Power Supply Power input: 100-240 V ~ 200 mA 50-60Hz 12-18VA; Power output: 12 VDC 0.5A
Assembled Dimensions
(W x D x H)
201 x 317 x 93 mm
7.9 x 12.5 x 3.7 inch
Communication RS232
Operating Temperature
Range
Operating conditions for ordinary lab application: +10°C to 30°C (operability guaranteed between
+5°C and 40°C).
Storage Temperature Range
Humidity: maximum relative humidity 80 % for temperatures up to 30°C,
decreasin
g
linearl
y
to 50% relative humidit
y
at 40°C
Storage Conditions -10°C to 60°C, humidity 10% to 90%, without condensation
Net Weight 7.7 lb / 3.5 kg
Shipping Weight 11 lb / 5 kg
Shipping Dimensions
(W x D x H)
550 x 385 x 291 mm
22 x 15 x 12 inch
*SRP refers to the standard deviation for n replicate weightings (n 10).