What does Err 9.5 mean on OHAUS Laboratory Equipment?
- BBrandy MendezAug 17, 2025
If your OHAUS Laboratory Equipment is displaying 'Err 9.5', the loadcell may not be working correctly.
What does Err 9.5 mean on OHAUS Laboratory Equipment?
If your OHAUS Laboratory Equipment is displaying 'Err 9.5', the loadcell may not be working correctly.
What does Err 53 mean on OHAUS Laboratory Equipment?
If your OHAUS Laboratory Equipment shows 'Err 53', it means there is an issue with the EEPROM.
Why does my OHAUS Laboratory Equipment show '--NO--' when taring?
The '--NO--' error on your OHAUS Laboratory Equipment indicates that the tare is out of range.
Why does my OHAUS Laboratory Equipment show '--NO--' when zeroing?
The '--NO--' error on your OHAUS Laboratory Equipment indicates that the zero is out of range.
What does 'LoW.rEF' mean on OHAUS Laboratory Equipment?
The 'LoW.rEF' error on your OHAUS Laboratory Equipment means that the reference weight is low, and this only applies in counting mode.
What does Err 8.1 mean on OHAUS Laboratory Equipment?
The 'Err 8.1' error on your OHAUS Laboratory Equipment indicates an over init zero range.
What does Err 8.2 mean on OHAUS Laboratory Equipment?
The 'Err 8.2' error on your OHAUS Laboratory Equipment indicates an under init zero range.
What does IDnr.Err mean on OHAUS Laboratory Equipment?
The 'IDnr.Err' error on your OHAUS Laboratory Equipment indicates an IDNR error.
What does PUSH.LFt mean on OHAUS Laboratory Equipment?
The 'PUSH.LFt' message on your OHAUS Laboratory Equipment indicates that you need to push the LFT button, specifically for PR.
What does 'No CAL' mean on OHAUS Laboratory Equipment?
The 'No CAL' error on your OHAUS Laboratory Equipment indicates a production calibration failure.
Capacity | 2200 g |
---|---|
Readability | 0.01 g |
Calibration | External |
Linearity | ±0.02 g |
Repeatability | 0.01 g |
Operating Temperature Range | 10 °C to 40 °C |
Display | LCD |
Weight | 1.5 kg |
Power Supply | AC Adapter (Included) |
Units of Measurement | g, oz |
Interface | RS-232 |
Essential safety guidelines for installation, connection, and operation to prevent injury or damage.
Steps for connecting the power supply and recommended acclimation periods.
Guide on connecting the balance to computers or printers via the RS232 port.
Importance and steps for calibrating the balance for accurate measurements.
Overview of internal, span, and linearity calibration procedures.
Setting legal for trade status and selecting measurement units.
Steps to configure, verify, and seal the balance for commercial applications.
Official verification process and methods for sealing the balance.
Procedures for periodic calibration verification and safe cleaning of the balance.
A table listing symptoms, causes, and solutions for balance errors.