Ranger™ 4000 Series
Instruction Manual

Other manuals for OHAUS RANGER 4000 Series

Questions and Answers

  • M
    Monica JimenezAug 20, 2025
    How to troubleshoot OHAUS RANGER 4000 that will not turn on?
    • J
      Jason HendersonAug 20, 2025
      If your OHAUS Scales will not turn on, ensure that the scale has power. Check the connections and verify the voltage.
  • R
    Ryan BriggsAug 22, 2025
    What to do if my OHAUS RANGER 4000 Scales show Err 8.3?
    • A
      Alexandra ChanAug 22, 2025
      If the OHAUS Scales display shows 'Err 8.3', it means the weight reading exceeds the overload limit. Reduce the weight on the scale to be within the specified range.
  • M
    Michael JohnsonAug 25, 2025
    What does Err 8.5 mean on OHAUS RANGER 4000 Scales and how to fix it?
    • J
      Jason RobertsonAug 25, 2025
      If the OHAUS Scales display shows 'Err 8.5', it means the tare value is out of range. Adjust the tare value to ensure it falls within the acceptable limits.
  • J
    jameshughesAug 26, 2025
    Why is the battery icon flashing on my OHAUS RANGER 4000 Scales?
    • B
      brookskylieAug 26, 2025
      If your OHAUS Scales show a flashing battery icon, it indicates a low battery. Connect the scale to AC power to charge the battery.
  • A
    Andrew MartinAug 28, 2025
    What does Err 8.6 mean on OHAUS Scales?
    • B
      Benjamin BuckleyAug 28, 2025
      If the OHAUS Scales display shows 'Err 8.6', it means the weight exceeds 6 digits. Reduce the weight on the scale to be within the display limit.
  • M
    Mark CurtisAug 29, 2025
    Why are my OHAUS RANGER 4000 Scales showing poor accuracy?
    • M
      michael18Aug 29, 2025
      If your OHAUS Scales are not accurate, it might be due to improper calibration or an unstable environment. Try performing a calibration procedure. If the issue persists, move the scale to a more stable location.
  • M
    Ms. Jessica FisherAug 31, 2025
    What does Err 8.4 mean on OHAUS RANGER 4000 Scales and how do I resolve it?
    • S
      stanleyandersonSep 1, 2025
      If the OHAUS Scales display shows 'Err 8.4', it indicates that the weight reading is below the underload limit. Increase the weight on the scale to be within the specified range.
  • D
    Dorothy JimenezSep 2, 2025
    What does Err 8.1 mean on OHAUS RANGER 4000 Scales?
    • S
      Sara SalazarSep 2, 2025
      If the OHAUS Scales display shows 'Err 8.1', it means the weight reading exceeds the Power On Zero limit. Ensure the weight is within the acceptable range when powering on the scale.
  • D
    daviddavisSep 4, 2025
    What does Err 8.2 mean on OHAUS RANGER 4000 Scales and how to fix it?
    • K
      kathleennicholsSep 4, 2025
      If the OHAUS Scales display shows 'Err 8.2', it means the weight reading is below the Power On Zero limit. Ensure the weight is within the acceptable range when powering on the scale.
  • A
    Amanda MartinezSep 6, 2025
    What does CAL E mean on OHAUS RANGER 4000 Scales?
    • M
      Matthew WalkerSep 6, 2025
      If your OHAUS Scales display shows 'CAL E', this indicates a calibration error. This may be caused by: * A calibration value outside of the allowed limits * An unstable environment * An incorrect calibration weight Move the scale to a more suitable location, and use the correct calibration weight.

Summary

Introduction

1.1 Signal Warnings and Symbols

Defines signal words and symbols used for safety notes and warnings.

1.2 Safety Precautions

Lists essential safety precautions for operating the scale.

Installation

2.4 Leveling Equipment

Explains how to use the level indicator and adjust feet for accurate weighing.

2.5 Connecting Power

Instructions for connecting the scale to AC power or battery.

Battery Power Operation

Operation

3.1 Controls Overview

Overview of the scale's control panel and button functions.

3.6 Check Mode

Instructions for comparing items against target weight ranges.

3.6.1 Check Weighing

Procedure for weight-based comparison within target ranges.

Menu Settings

4.2 Calibration Menu

Procedure for performing scale calibration to ensure accuracy.

4.3 Setup Menu

Lists and explains various scale parameters that can be configured.

Legal for Trade

5.2 Settings for Verification

Steps for verification and sealing, including LFT switch settings.

5.3 Verification and Sealing

Overview of the verification and sealing process by officials.

Maintenance

6.1 Cleaning

Safety warnings and instructions for cleaning the scale.

6.2 Troubleshooting

Lists common problems, causes, and remedies for scale issues.

OHAUS RANGER 4000 Series Specifications

General IconGeneral
Readability0.1 g
DisplayLCD with backlight
Power SupplyAC adapter
Battery Life100 hours
Legal for TradeYes
ProtectionIP54
Operating Temperature32°F to 104°F (0°C to 40°C)
Relative Humidity10%
InterfaceRS232
Units of Measurementg, kg, lb, oz

Related product manuals