How to troubleshoot OHAUS RANGER 4000 that will not turn on?
- JJason HendersonAug 20, 2025
If your OHAUS Scales will not turn on, ensure that the scale has power. Check the connections and verify the voltage.
How to troubleshoot OHAUS RANGER 4000 that will not turn on?
If your OHAUS Scales will not turn on, ensure that the scale has power. Check the connections and verify the voltage.
What to do if my OHAUS RANGER 4000 Scales show Err 8.3?
If the OHAUS Scales display shows 'Err 8.3', it means the weight reading exceeds the overload limit. Reduce the weight on the scale to be within the specified range.
What does Err 8.5 mean on OHAUS RANGER 4000 Scales and how to fix it?
If the OHAUS Scales display shows 'Err 8.5', it means the tare value is out of range. Adjust the tare value to ensure it falls within the acceptable limits.
Why is the battery icon flashing on my OHAUS RANGER 4000 Scales?
If your OHAUS Scales show a flashing battery icon, it indicates a low battery. Connect the scale to AC power to charge the battery.
What does Err 8.6 mean on OHAUS Scales?
If the OHAUS Scales display shows 'Err 8.6', it means the weight exceeds 6 digits. Reduce the weight on the scale to be within the display limit.
Why are my OHAUS RANGER 4000 Scales showing poor accuracy?
If your OHAUS Scales are not accurate, it might be due to improper calibration or an unstable environment. Try performing a calibration procedure. If the issue persists, move the scale to a more stable location.
What does Err 8.4 mean on OHAUS RANGER 4000 Scales and how do I resolve it?
If the OHAUS Scales display shows 'Err 8.4', it indicates that the weight reading is below the underload limit. Increase the weight on the scale to be within the specified range.
What does Err 8.1 mean on OHAUS RANGER 4000 Scales?
If the OHAUS Scales display shows 'Err 8.1', it means the weight reading exceeds the Power On Zero limit. Ensure the weight is within the acceptable range when powering on the scale.
What does Err 8.2 mean on OHAUS RANGER 4000 Scales and how to fix it?
If the OHAUS Scales display shows 'Err 8.2', it means the weight reading is below the Power On Zero limit. Ensure the weight is within the acceptable range when powering on the scale.
What does CAL E mean on OHAUS RANGER 4000 Scales?
If your OHAUS Scales display shows 'CAL E', this indicates a calibration error. This may be caused by: * A calibration value outside of the allowed limits * An unstable environment * An incorrect calibration weight Move the scale to a more suitable location, and use the correct calibration weight.
Readability | 0.1 g |
---|---|
Display | LCD with backlight |
Power Supply | AC adapter |
Battery Life | 100 hours |
Legal for Trade | Yes |
Protection | IP54 |
Operating Temperature | 32°F to 104°F (0°C to 40°C) |
Relative Humidity | 10% |
Interface | RS232 |
Units of Measurement | g, kg, lb, oz |