5/Disassembly
and
Repair
С.
Вей
Platform
Disassembly
(Figure
5-12)
1.
Remove
the
mattress
and
the
Plexiglas
cover.
2.
Remove
the
four
side
panels
from
the
bed
plat-
form.
3.
Remove
the
four
comer
blocks
from
the
bed
plat-
form.
4.
Use
a
1/2
inch
wrench
and
7/16
inch
wrench
to
remove
the
hydraulic
system
mounting
nut
and
stud
from
the
lower
support.
5.
Use
a
No.
2
Phillips
screw
driver
and
remove
the
four
retaining
rods
and
hooks
from
the
bottom
of
the
bed
platform.
6.
Slide
the
two
bed
pivot
rods
out
from each
side.
7.
Lift
the
bed
platform
off
the
lower
support.
D.
Bed
Platform
Assembly
(Figure
5-12)
1.
Place
the
bed
platform
in
position
on
the
lower
support.
2.
Insert
the
two
bed
pivot
rods
into
position
on
the
lower
support.
The
notch
must
face
the
bed
plat-
form
(upwards)
and
be
positioned
between
the
notches
in
the
bed
platform.
3.
Use
a
No.
2
Phillips
screw
driver
and
install
the
four
retaining
rods
and
hooks
on
the
bottom
of
the
bed
platform.
The
open
end
of
the.hooks
should
face
away
from
the
bed
platform.
4.
Use
a
1/2
inch
wrench
and
7/16
inch
wrench
to
install
the
hydraulic
system
mounting
nut
and
stud
on
the
lower
support.
5.
Replace
the
four
corner blocks
on
the
bed
plat-
form.
6.
Replace
the
four
side
panels
on
the
bed
platform.
7.
Replace
the
Plexiglas
cover.
8.
Replace
the
mattress.
E.
Hydraulic
System
Removal
(Figures
5-13
and
5-14)
Note:
Ohmeda
recommends
replacing
the
hydraulic
system
as
an
assembly.
The
system
uses
a
standard
synthetic hydraulic
oil.
Note:
The
unit
may
be
carefully
placed
on
it's
right
side
(looking
from
the
front)
or
the
bed
may
be
re-
moved
for
replacement
of
the
hydraulic
system.
1.
Use
a
1/2
inch
wrench
and
7/16
inch
wrench
to
remove
the
hydraulic
system
mounting
nut
and
stud
from
the
lower
support.
2.
Remove
the
two
Phillips
head
mounting
screws
which
hold
the
outer
(triangular
shaped)
cover
plate
in
position.
3.
Remove
the
four
Phillips
head
mounting
screws
which
hold
the
inner
(square
shaped)
cover
plate
in
position.
Take
care
to
ensure
that
the
tension
on
the
spring
is
released
carefully.
Note:
The
tilt
lever,
rod,
spring,
and
mounting
pin
can
be
removed
for
replacement
if
necessary.
Re-
move
the pin
from
the
top
of
the
bed
to
remove
the
spring
and
rod.
4.
Note
how
the
tubing
is
installed
in
parallel
and
does
not
overlap
until
it
reaches
the
storage
area.
5.
Remove
the
hydraulic
system
assembly
for
re-
placement.
F.
Hydrualic
System
Installation
(Figures
5-13
and
5-14)
1.
Transfer
the
mounting
pin
from
the
old
hydraulic
cylinder
to
the
new
hydraulic
cylinder.
2.
Place
the
hydraulic
cylinder
with
pin
in
position.
3.
Install
the
tubing
for
the
hydraulic
system
in
parallel
and
make
sure
it
does
not
overlap
until
it
reaches
the
storage
area.
Note:
The
tubing
must
not
be
stretched,
pinched
or
kinked
during
reassembly.
If
the
tubing
is
pinched
the
bed
will
not
tilt.
4.
Place
the
inner
(square
shaped)
cover
plate
in
position.
5.
Install
the
two
mounting
screws
closest
to
the
tilt
lever,
6.
Install
the
two
mounting
screws
closest
to
the
cylinder
at
the
edge
of
the
cover
plate.
7.
Coil
the
tubing
so
it
fits
in
the
triangle
area.
8.
Place
the
outer
(triangular
shaped)
cover
plate
in
position
and
replace
the
two
Phillips
head
mount-
ing
screws.