4/Calibration
and
Adjustments
CAUTION:
Use
the
Static
Control
Work
Station
(Part
No.
0175-2311-000)
to
help
ensure
that
static
charges
are
safely
conducted
to
ground.
The
Velos-
tat
material
is
conductive.
Do
not
place
electrically
powered
circuit
boards
on
it.
Note:
The
audio
alann
will
sound
for
about
2
sec-
onds
whenever
powering
up
the
unit.
Note:
Warm
up
the
unit
for
5
minutes
before
mak-
ing
these
adjustments.
A.
Control
Unit
Access
1..
Disconnect
the
power
cord
for
the
Infant
Warmer
System
from
the
wall
outlet.
2.
Remove
the
mounting
screws
for
the
back
panel.
B.
Power
Supply
Board Voltage
Checks
1.
Set
the
test
switch
(S1)
located
on
the
control
board
to
the
following
test
positions:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On")
2.
Connect
the
controller
assembly
to
the
power
source
(listed
voltage
+
10%),
and
switch
the
In-
fant
Warmer
“On”.
3.
Check
that
all
display
segments
are
“On”,
the
observation
lamp
is
“On”,
the
alarm
light
is
“On”,
the
heater
radiates
heat,
and-a
continuous
alarm
tone
sounds.
4.
Check
that
the
following
D.C.
voltages
are
pre-
sent
at
the
test
connector
(T1)
located
on
the
control
board.
Voltages
should
be
within
the
to-
lerances
specified:
TP7
Ground
(common)
TP3
(9.0
Vdc
ST.)
+9V
+
0.2V
(adjust
R4 on
P.S.
Bd.)
TP4
(5.0
Vde
LEDS)
+5V
+
0.2V
(replace
P.S.
Bd.)
TP5
(5.0
Vdc
ST.)
+5V
+
0.2V
(replace
P.S.
Bd.)
C.
Display
Brightness
Check
Note:
The
display
brightness
is
precalibrated
at
the
factory
and
should
only
require
adjustment
if
replac-
ing
a
component
on
the
display
board.
With
the
test
switches
set
as
follows:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On”)
Check
that
all
the
displays
are
lit
and
are
of
uniform
brightness.
If
the
displays
are
acceptable
proceed
to
Section
D.
If
the
displays
are
not
illuminated
adequately
proceed
with
the
adjustment
procedure.
Adjustment
Procedure
CAUTION:
The
back
panel
and
display
panel
may
drop
down
when
the
bottom
cover
mounting
screws
are
removed,
Be
sure
to
secure
the
panels
with
tape
before
disassembly.
1.
Set
the
test
switch
($1)
located
on
the
control
board
to
the
following
test
positions:
Switch
#1,#2,#4
Ореп
("ОН")
Switch
#3
Closed
(“On")
2.
Tape
the
display
panel and back
panel
to
the
top
cover.
3.
Remove
the
four
outside
corner
mounting
screws
from
the
bottom
of
the
display
panel.
Do
not
re-
move
the
two
inside
mounting
screws.
4.
Connect
a
digital
voltmeter
across
R10 located
on
the
bottom
edge
of
the
display
board.
5.
Adjust
R9 on
the
display
board
until
the
voltage
across
R10
is
3.30V
+
0.2V.
6.
Verify
that
all
segments
of
all
displays
are
lit
and
are
of
uniform
brightness.
D.
Alarm
Volume
Adjustment
Note:
The
alarm
volume
and
frequency
are
precalib-
rated
at
the
factory
and
should
only
require
adjust-
ment
if
replacing
a
component
on
the
control
board.
1.
With
the
test
switches
set
as
follows:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On”)
2.
Ensure
the
alarm
tone
and
volume
are
adequate.
Verify
that
the
audio
alarm
level
is
adequate
ina
lecation
with
a
background
noise
level
of
55
dBA
max.
If
the
audio
alarm
level
is
acceptable
go
to
step
E.
If
the
audio
alarm
level
is
unacceptable
proceed
with
the
adjustment
procedure.
Adjustment
Procedure
3.
Verify
the
frequency
output
at
U7
pin
9
is
2
kHz
+
0.1
kHz.
Adjust
R38
on
the
control
board
as
re-
quired.
Note:
If
test
equipment
is
not
available
for
check-
ing
the
2
kHz
frequency,
adjust
R38
for
maximum
sound
level.
4.
Verify
that
R37
on
the
control
board
is
set
fully
CCW
(maximum
volume).
E.
Analog
to
Digital
Converter
(ADC)
Note:
The
following
resistance
values
are
available
on
the
Temperature
Simulator
and
the
switch
posi-
tions
are
listed
in
parentheses.
1.
Switch
the
power
switch
“Off”.
2.
Set
the
control
board
test
DIP
switch
(located
on
the
upper
edge)
to
the
calibration
positions
as
follows:
Switch.#1,
#2,
#3
Open
(“Off”)
Switch
#4
Closed
(“On”)
3.
Connect
a
resistance
of
5900
ohms
+
0.1%
(17)
to
the
patient
jack
connector.
4-1