4 - Insert the part back into the
OLCT100 and then mount the
cable gland (not supplied).
Cable connection
The connection of the cable between the detector and controller should be made
w ith the pow er off. The site should be at equal potential
Connect the cable to the detector side before connecting the controller side.
After the wiring has been completed, connect the cable screen to the ground
terminal of the controller.
022
Figure 8: connection for a 2-w ire 4-20 Ma detector