EasyManua.ls Logo

Oldham OX 2000

Default Icon
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
3
3
.
.
C
C
A
A
L
L
I
I
B
B
R
R
A
A
T
T
I
I
O
O
N
N
The only means of checking that the instrument’s detection function is working
properly is the calibration performed with the help of a standard gas.
Calibration must be done :
- At least once a year following the use and the type of sensor
- whenever the instrument has been exposed to overly high concentrations of
gas
- when an instrument has been stored for more than one month and without
use
- when changing of sensors
- when fault appears
Within the calibration the menu "visible and audible" alarms are not triggered.
3.1. Calibration procedure
Calibration ” menu consists of
- setting the zero in clean air and the sensitivity with calibration gas in
TX2000(+) version (toxic gases).
- Adjust oxygen sensitivity in OX 2000(+) version.
3.2. Adjusting the zero in clean air (only for toximeter version)
Display the “ programming ” menu
To display the "calibration" menu
Display the"calibration" menu
To enter the “ calibration ”menu
PrG
CAL
ENTER

Other manuals for Oldham OX 2000

Related product manuals