The only means of checking that the instrument’s detection function is working
properly is the calibration performed with the help of a standard gas.
Calibration must be done :
- At least once a year following the use and the type of sensor
- whenever the instrument has been exposed to overly high concentrations of
gas
- when an instrument has been stored for more than one month and without
use
- when changing of sensors
- when fault appears
Within the calibration the menu "visible and audible" alarms are not triggered.
3.1. Calibration procedure
Calibration ” menu consists of
- setting the zero in clean air and the sensitivity with calibration gas in
TX2000(+) version (toxic gases).
- Adjust oxygen sensitivity in OX 2000(+) version.
3.2. Adjusting the zero in clean air (only for toximeter version)
Display the “ programming ” menu
To display the "calibration" menu
Display the"calibration" menu
To enter the “ calibration ”menu
PrG
CAL
ENTER

Other manuals for Oldham OX 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Oldham OX 2000 and is the answer not in the manual?

Oldham OX 2000 Specifications

General IconGeneral
BrandOldham
ModelOX 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals