3.2 Manual Control
!
Sol Time- Opening time for acetonitrile valve, here 800 ms. It can be
lowered by pushing the button (-50 ms)
Sol Time+ Opening time can be adjusted higher by pushing the button (+50 ms).
Run Sol A program that fills all the columns with the desired amount of
acetonitrile, following with activating of Pump Run. You can check
whether all columns have been filled with solvent and subsequently sucked
dry. Should be performed prior to every run.
Pump Sucktion time, here 2000 ms. It can be lowered by pushing the button
Time- (-50ms)
Pump Sucktion time can be higher by pushing the button (+50 ms).
Time+
Run Activate the pump.
Pump
Zero Set Reaction Block in center-position
Position
Sol Air Switch On and Off the valve from the nitrogen cylinder to acetonitrile (Sol)
bottle