102
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
7.5 Manual cleaning
After completing the leakage test, perform manual cleaning according to the
procedures described below.
In case of excessive bleeding and/or delayed reprocessing, perform “Presoak for
excessive bleeding and/or delayed reprocessing” on page 113 and follow the
procedure described below.
Reusable part that is normally reprocessed with the
endoscope
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
Water-resistant cap (2 pcs) (MH-553)
Chain for water-resistance caps (MAJ-1723)
Personal protective equipment
Soft brush
•30cm
3
(30 ml) syringe
Clean, lint-free cloths
Large basins
Detergent solution
Clean water (see Section 6.2)
Suction cleaning adapter (MAJ-222)
Channel cleaning brush or
single-use channel cleaning brush
(BW-15B)
(BW-201B)
Channel-opening cleaning brush or
single use channel-opening cleaning
brush
(MH-507)
(MAJ-1339)
Channel cleaning brush (BW-7B)
Suction pump
Single use combination cleaning brush (BW-411B)
Single use single-ended cleaning brush (BW-400B)