Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
131
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
7.10 Care of the ultrasonic cable (MAJ-1597,
MAJ-1722)
Equipment needed
Care
1.
If the cable is soiled with blood or other potentially infectious materials, wipe
off all gross debris using a soft, lint-free cloth moistened with neutral
detergent.
2. To remove dust, dirt and other non-patient debris, wipe the cable’s surface
with a soft, lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
3. Make sure that the cable is completely dry.
Never apply water to the ultrasonic cable. Equipment
damage can result.
The ultrasonic cable is not waterproof. Never immerse it in
disinfectant solution or any other fluids.
Alcohol is not a sterilant or high-level disinfectant.
Clean, lint-free cloth
Neutral detergent
70% ethyl or isopropyl alcohol

Table of Contents

Other manuals for Olympus BF-UC180F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus BF-UC180F and is the answer not in the manual?

Olympus BF-UC180F Specifications

General IconGeneral
ModelBF-UC180F
Insertion Tube Diameter6.0 mm
Working Length600 mm
Field of View120°
Distal End Angulation Up180°
Distal End Angulation Down130°
Light SourceXenon
Video OutputHDTV
Instrument Channel Diameter2.0 mm
Direction of ViewForward
Depth of Field3-50 mm

Related product manuals