Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
97
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Figure 7.16
5. Disconnect the ultrasonic cable from the endoscope connector.
6. Disconnect the videoscope cable from the endoscope connector.
7. Disconnect the endoscope connector from the light source.
8. Transport the endoscope and the container holding the other items to the
reprocessing area.
7.4 Leakage testing
After precleaning, perform leakage testing on the endoscope to ensure that it is
waterproof.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
Personal protective equipment
Large basin
Clean water (see Section 6.2)
Maintenance unit or light source (MU-1 or CLV-180, CLV-160, CLV-U40)
Leakage tester (MB-155)
Water-resistant cap (2 pcs) (MH-553)

Table of Contents

Other manuals for Olympus BF-UC180F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus BF-UC180F and is the answer not in the manual?

Olympus BF-UC180F Specifications

General IconGeneral
ModelBF-UC180F
Insertion Tube Diameter6.0 mm
Working Length600 mm
Field of View120°
Distal End Angulation Up180°
Distal End Angulation Down130°
Light SourceXenon
Video OutputHDTV
Instrument Channel Diameter2.0 mm
Direction of ViewForward
Depth of Field3-50 mm

Related product manuals