Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Assembling the suction valve (MAJ-207)
1.
Press the valve into the main body (see Figure 7.26).
2. Attach the spring to the main body (see Figure 7.26).
3. Insert the piston through the components and press it down a few times
(see Figure 7.26).
Figure 7.26
Piston
3.
Spring
2.
Main body
1.
Valve
Press down a
few times.