Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
109
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Aspirating detergent solution into the instrument/suction
channels
1.
Attach the suction cleaning adapter’s port cap to the instrument channel
port. Attach the suction cylinder cap to the suction cylinder (see Figure
7.21).
2. Connect the suction tube from the suction pump to the suction opening of
the suction cleaning adapter.
3. Turn the suction pump ON.
4. Immerse the endoscope’s distal end in the detergent solution.
5. Aspirate the detergent solution for approximately 30 seconds.
6. Turn the suction pump OFF.
7. Disconnect the suction tube from the suction cleaning adapter.