Appendix D: Acoustic Output Information Accordance with IEC 60601-2-37
169
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
SSD-5
Transducer Model:
BF-UC180F
Operating Mode:
Mflow-mode
Table 101 – Acoustic output reporting table
Index label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1cm
2
A
aprt
>1cm
2
Maximum Index Value 1.6 – 0.42 – 1.3 0.78
Associated
acoustic
parameters
p
r,
(MPa) 3.6
P (mW) –15 13 15
Min of [P
(z
s
), I
ta,
(z
s
)]
(mW)
–
z
s
(cm) –
z
bp
(cm) –
z
b
(cm) 1.1
z at max. I
pi,
(cm) 0.33
d
eq
(z
b
)
(cm)
0.14
ƒ
awf
(MHz) 5.1 – 6.1 – 6.0 6.0
Dim of A
aprt
X(cm) – 0.67 – 0.67 0.67
Y(cm)
– 0.26 – 0.26 0.26
Other
information
t
d
(µsec) 0.75
prr (Hz) 1090
p
r
at max. I
pi
(MPa) 3.8
d
eq
at max. I
pi
(cm) 0.14
I
pa,
at max.MI
(W/cm
2
)
535
Operating
control
conditions
DVA% (%) 89 – 73 – 92 55
Function – – – – – –
B/W Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
B/W Focus n/a n/a n/a n/a n/a n/a
B/W-Range (cm) 2.0 – 2.0 – 2.0 2.0
B/W Line Density n/a n/a n/a n/a n/a n/a
B/W Scan Area (%) n/a n/a n/a n/a n/a n/a
Doppler Image Select n/a n/a n/a n/a n/a n/a
Doppler Frequency Select (MHz) n/a n/a n/a n/a n/a n/a
Doppler Focus n/a n/a n/a n/a n/a n/a
Doppler Velocity Range (cm/sec) n/a n/a n/a n/a n/a n/a
Color Image Select PENET – PENET – PENET PENET
Color Doppler Frequency
Select
(MHz) 5MHz – 6MHz – 6MHz 6MHz
Flow Focus F1 – F5 – F3 F7
Color Velocity Range (m/sec) 0.08 – 0.17 – 0.10 0.40
Color Line Density n/a n/a n/a n/a n/a n/a
Color Average n/a n/a n/a n/a n/a n/a
Flow Filter n/a n/a n/a n/a n/a n/a
Flow Scan Area (%) n/a n/a n/a n/a n/a n/a
Triplex Velocity Range n/a n/a n/a n/a n/a n/a
NOTE1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 Information on MI and TI need not be provided if the equipment meets both exemption clauses given in 51.2 aa) and 51.2
dd).