Appendix D: Acoustic Output Information Accordance with IEC 60601-2-37
173
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
SSD-5
Transducer Model:
BF-UC180F
Operating Mode:
B/Bflow/D-mode
Table 101 – Acoustic output reporting table
Index label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1cm
2
A
aprt
>1cm
2
Maximum Index Value 1.6 – 0.24 – 1.2 0.84
Associated
acoustic
parameters
p
r,
(MPa) 3.9
P (mW) –8.5 12 15
Min of [P
(z
s
), I
ta,
(z
s
)]
(mW)
–
z
s
(cm) –
z
bp
(cm) –
z
b
(cm) 1.1
z at max. I
pi,
(cm) 0.35
d
eq
(z
b
)
(cm)
0.15
ƒ
awf
(MHz) 6.0 – 6.1 – 6.0 5.9, 6.0
Dim of A
aprt
X(cm) – 0.67 – 0.67 0.67
Y(cm)
– 0.26 – 0.26 0.26
Other
information
t
d
(µsec) 0.58
prr (Hz) 2780
p
r
at max. I
pi
(MPa) 4.1
d
eq
at max. I
pi
(cm) 0.14
I
pa,
at max.MI
(W/cm
2
)
657
Operating
control
conditions
DVA% (%) 76 – 65 – 65 71
Function – – – – – –
B/W Frequency Select (MHz) 7.5MHz – 7.5MHz – 12MHz 5MHz
B/W Focus F3 – F7 – F1 F7
B/W-Range (cm) 10.0 – 2.0 – 2.0 2.0
B/W Line Density HIGH – HIGH – HIGH HIGH
B/W Scan Area (%) 25 – 25 – 25 25
Doppler Image Select PENET – PENET – PENET PENET
Doppler Frequency Select (MHz) 6MHz – 6MHz – 6MHz 6MHz
Doppler Focus F4 – F5 – F3 F8
Doppler Velocity Range (cm/sec) 16.6 – 12.4 – 24.9 12.4
Color Image Select PENET – PENET – PENET PENET
Color Doppler Frequency
Select
(MHz) 6MHz – 6MHz – 6MHz 6MHz
Flow Focus F1 – F5 – F6 F1
Color Velocity Range (m/sec) 0.08 – 0.06 – 0.12 0.06
Color Line Density LOW – LOW – LOW LOW
Color Average HIGH – HIGH – HIGH HIGH
Flow Filter #4 – #4 – #4 #4
Flow Scan Area (%) 5 – 5 – 5 5
Triplex Velocity Range HIGH – HIGH – HIGH HIGH
NOTE1 Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
NOTE2 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or
neonatal cephalic uses.
NOTE3 Information on MI and TI need not be provided if the equipment meets both exemption clauses given in 51.2 aa) and 51.2
dd).