66
Chapter 4 Operation
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Do not open the tip of the EndoTherapy accessory or extend
the tip of the EndoTherapy accessory from its sheath while
inserting the EndoTherapy accessory into the instrument
channel. The instrument channel and/or the EndoTherapy
accessory may become damaged.
4. Hold the EndoTherapy accessory approximately 4 cm from the biopsy valve
and advance it slowly and straight into the valve using slow, short strokes
while observing the endoscopic image.
Operation of EndoTherapy accessories
Operate the EndoTherapy accessory according to the directions given in its
instruction manual.
Withdrawal of EndoTherapy accessories
Do not withdraw the EndoTherapy accessory if the tip is open
or extended from its sheath; patient injury and/or instrument
damage may occur.
If the EndoTherapy accessory cannot be withdrawn from the
endoscope, carefully withdraw both the endoscope and the
EndoTherapy accessory together under endoscopic
observation. Take care not to cause tissue trauma.
Withdraw the EndoTherapy accessory slowly while the tip of the EndoTherapy
accessory is closed and/or retracted into its sheath.