EasyManua.ls Logo

Olympus bx40 - Page 40

Olympus bx40
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aperture
iris
diaphragm
Objective
pupil
|
Fig.
54
(2)
1
lara
tasa
ας
«ελ
N.A.
scale
i
PM
|
Fig.
55
(Figs.
54,
55)
|
¢
The
aperture
iris
diaphragm
determines
the
numerical
aperture
of
the
illumination
system.
Matching
the
numerical
aperture
of
the
illumination
system
with
that
of
the
objective
provides
better
image
resolution
and
contrast,
and
also
increases
the
depth
of
focus.
|
Aperture
Iris
Diaphragm
*
Since
the
contrast
of
microscopic
specimens
is
ordinarily
low,
setting
the
condenser
aperture
iris
diaphragm
to
70~80%
of
the
N.A.
of
the
objective
in
use
is
usually
recommended.
When
necessary,
adjust
this
ratio
by
removing
the
eyepiece
and
peeking
into
the
eyepiece
sleeve
to
see
the
image
shown
in
Fig.
54.
E
Using
the
Numerical
Aperture
Scale
Al
Set
the
condenser
numerical
aperture
to
about
80%
of
the
NA
value
@
indicated
on
the
objective.
(Fig.
55)
Example:
With
the
Plan
40X
(NA
0.65),
set
the
scale
to
0.65
x
0.8
=
0.5.
Compatibility
of
Objectives
and
Condensers
ns
E
q
386
A
DE
ως
Usable
by
moving
Usable
5
front
element
out
of
Usable
to
FN22
the
light
path*?
Usable
Usable
eme
NA
not
fully
adequate*'
*'
When
using
the
U-SC
swing-out
achromat
condenser
together
with
the
2X
or
4X
objective,
fully
open
the
condenser
aperture
and
use
the
field
iris
diaphragm
in
the
base
as
aperture
diaphram.
*2
Although
slightly
inadequate
NA
results
in
a
somewhat
darker
field
of
view
with
a
100X
objective,
the
combination
is
usable.
©
To
obtain
better
illumination,
use
of
the
U-ULC
is
recommended
in
photomicrography
when
using
the
2X
or
4X
objective.
Table
4
A-18
A0

Other manuals for Olympus bx40

Related product manuals