32
Proper selection of the power supply cord
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to
“Specifications” and “Certified Cord” below:
Caution : In case you use a non-approved power supply cord for Olympus products, Olympus can no
longer warrant the electrical safety of the equipment.
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked
with an agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of the
agencies listed in Table 1. In case you are unable to buy locally the power supply cord which is approved by one
of the agencies mentioned in Table 1, please use replacements approved by any other equivalent and authorized
agencies in your country.
Table 1 Certified cord
Specifications
Voltage rating
Current rating
Temperature rating
Length
Fittings configuration
125 V AC (for 100-120 V AC area) or, 250 V AC (for 220-240 V AC area)
6 A minimum
60 ˚C minimum
3.05 m maximum
Grounding type attachment plug cap. Opposite terminates in molded-on
IEC configuration appliance coupling.