28
Problem Cause Remedy Page
The picture is not in focus. The microscope is not focused
properly.
Adjust the focus correctly with the fine
adjustment knob.
The parfocality is not adjusted
properly.
Adjust parfocality with the camera
adapter.
The aperture iris diaphragm of the
condenser is open too wide.
Close the aperture iris diaphragm a
little.
The field iris diaphragm is not set
properly.
Adjust the field iris diaphragm until the
image circumscribes the field of view.
Lens components of the microscope
are contaminated or the cover glass
on the front of the camera is stained.
Clean the objective, photography lens,
condenser and/or window lens of the
microscope, or clean the cover glass
on the bottom of the camera head.
6
The microscope and/or camera
are subjected to vibration during
recording.
Record images in an environment in
which the microscope and camera
are not vibrated. It is effective to use an
anti-vibration bench.
The DP2-TWAIN or cellSens /
OLYMPUS Stream window is not
displayed correctly or the menu
characters are not displayed correctly.
The resolution setting of the screen is
incorrect.
Set the resolution setting at 1280 x
1024 or more in the properties of the
screen.
The large font has been selected for
the font size of the screen.
Select a small font in the properties of
the screen.
Another 1394 device cannot be used
simultaneously.
The DP26-dedicated driver is installed. When the DP26 is connected to a PC,
other 1394 devices cannot be used on
the same PC.