11
DP22/DP27
15. The image of dark specimen under the fluorescence observation or the darkfield observation (specimen that
needs exposure of 1/2 sec. or more at ISO 100 equivalent) cannot be acquired. When acquiring the image of
the particularly dark specimen, the lack of pixels caused by the cosmic ray or the constant bright points may be
noticeable, but they are not phenomena caused by the failure of the equipment.
16.
If the illumination light of the micr
oscope is set too bright, the color unevenness may be generated in the acquired
image. In such a case, adjust the light intensity of the illumination to an appropriate level (for example, by
lowering the lamp voltage or by inserting an attenuation filter). Or use the shading correction function. For details
of the shading correction function, refer to the online help and the instruction manual according to the relevant
connection status.
17.
When combining this equipment with CX series or CKX series micr
oscope, set the light intensity of the illumination
closer to the maximum level in order to gain the best color reproduction.
Recommended specifications of the monitor
Monitor available for full color view of 1280 x 1024 or more.