100
Chapter 4 Operation
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
Operating the zoom function (for GIF-H260Z, CF-H260AZL/I,
PCF-Q260AZI only)
1.
Refer to the magnification controller’s instruction manual for the operation of
the magnification controller (for CF-H260AZL/I only).
2. When magnifying the endoscopic image, move the zoom lever toward “ T”
and move the distal end of the endoscope closer to the object you wish to
observe.
• While the endoscopic image is magnified, objects at a
distance from the distal end of the endoscope may be out of
focus. This is not a malfunction; move the distal end closer to
the object you wish to observe.
• The CV-260SL, CV-260 indicates the “scale display” on a
monitor. Refer to the instruction manual for the CV-260SL,
CV-260 for details of the “scale display” function.
Operating the endoscope position detecting function (for
CF-Q260DL/I, CF-H260DL/I only)
Refer to the endoscope position detecting unit’s instruction manual for
instructions on how to operate the endoscope position detecting unit.
4.2 Using EndoTherapy accessories
For more information on combining the endoscope with particular EndoTherapy
accessories, refer to the “System chart” in the Appendix and the instruction
manuals for the accessories.
• Do not use the GIF-N260, GIF-XP260 for high-frequency
cauterization or laser cauterization treatment. Otherwise,
patient injury or equipment damage may result.