Chapter 3 Preparation and Inspection
73
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
3.4 Attaching accessories to the endoscope
The air/water valve and the suction valve do not require
lubrication. Lubricants can cause swelling of the valves’
seals, which will impair valve function.
Attaching the suction valve
1.
Align the two metal ridges on the underside of the suction valve with the two
holes in the suction cylinder.
2. Attach the suction valve to the suction cylinder of the endoscope (see
Figures 3.16 and 3.17). Confirm that the valve fits properly without any
bulging of the skirt. Also confirm that the valve cannot be rotated.
Figure 3.16
The suction valve will make a whistling noise when it is dry;
this does not indicate a malfunction.
Side view
Bottom view Top view
Skirt
Two metal ridges
Suction cylinder
Suction cylinder
Two holes