102
Chapter 4 Operation
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
When using a distal attachment, the distal end of the
endoscope becomes longer and its outer diameter is larger.
Handle the endoscope carefully so as not to cause
perforation or other patient injury. When performing
endoscopic treatment using this equipment, take extra care.
When the distal attachment is mounted on the endoscope, do
not angulate the endoscope abruptly. This could cause
patient injury, such as mucous membrane damage.
Do not inflate air or a nonflammable gas excessively into the
patient. This could cause gas embolism.
When using a biopsy forceps with a needle, confirm that the
needle is not excessively bent. A bent needle could protrude
from the closed cups of the biopsy forceps. Using biopsy
forceps with a protruding needle could damage the
instrument channel and/or cause patient injury.
When using an injector, be sure not to extend or retract the
needle from the catheter of the injector until the injector is
extended from the distal end of the endoscope. The needle
could damage the instrument channel if extended inside the
channel, or if the injector is inserted or withdrawn while the
needle is extended.
Insertion of EndoTherapy accessories into the endoscope
Do not insert EndoTherapy accessories forcibly or abruptly.
Otherwise, the EndoTherapy accessory may extend from the
distal end of the endoscope abruptly, which could cause
patient injury, bleeding, and/or perforation.
When the biopsy valve’s cap is detached from the main body,
it is easier to insert an EndoTherapy accessory into the
instrument channel port (see Figure 3.12 on page 70).
However, the open biopsy valve, after withdrawing an
EndoTherapy accessory, can reduce the efficacy of the
endoscope’s suction system and may leak or spray patient
debris or fluids, posing an infection control risk. When not
using an EndoTherapy accessory, attach the cap to the main
body of the biopsy valve.