Chapter 5 Troubleshooting
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
Auxiliary water feeding (for endoscopes with auxiliary
water feeding only)
EndoTherapy accessories
Endoscope position detecting (for CF-Q260DL/I,
CF-H260DL/I only)
Other
Irregularity
description
Possible cause Solution
The auxiliary water
inlet cap is leaking.
The auxiliary water inlet cap
is worn out.
Replace it with a new one.
The auxiliary water inlet cap
is incorrectly installed.
Install the auxiliary water inlet cap
correctly.
No water feeding. The auxiliary water tube is
damaged.
Replace it with a new one.
Irregularity
description
Possible cause Solution
EndoTherapy
accessory does not
pass through the
instrument channel
smoothly.
An incompatible
EndoTherapy accessory is
being used.
Refer to the “System chart” in the
Appendix and select a compatible
EndoTherapy accessory. Confirm
that the color code on the
EndoTherapy accessory matches
that on the endoscope.
Irregularity
description
Possible cause Solution
The scope model
is not displayed.
The position detecting unit
is OFF.
Turn ON the position detecting unit.
The UPD cable is not
connected properly.
Connect the UPD cable correctly.
Irregularity
description
Possible cause Solution
The remote switch
does not work.
The wrong remote switch is
operated.
Operate the correct remote switch.
The remote switch function
has been set incorrectly.
Set the remote switch function
correctly as described in the video
system center’s instruction manual.