78
Chapter 3 Preparation and Inspection
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
10. Align the notch on the endoscope plug of the magnification control cable
with mark 3 on the zoom connector of the endoscope and push the
endoscope plug into the zoom connector until it stops (for CF-H260AZL/I
only, see Figure 3.22).
Figure 3.22
If force is used on the magnification control cable, the
endoscope is designed to detach at the endoscope plug, not
at the endoscope connector (for CF-H260AZL/I only).
11. Align the notch on the position detecting connector of the UPD cable with
mark 3 on the UPD scope connector of the endoscope and push the
position detecting connector into the UPD scope connector until it stops (for
CF-Q260DL/I, CF-H260DL/I only, see Figure 3.23).
Figure 3.23
If force is used on the UPD cable, the endoscope is designed
to detach at the position detecting connector, not at the
endoscope connector (for CF-Q260DL/I, CF-H260DL/I only).
Notch
Endoscope plug
Mark 3
(protrusion)
Notch
Mark 3
(protrusion)
Position detecting connector