•
Log on to with the Username "ALMF".
• Start the microscope control software using the icon on the desktop
Switching off
•
Check if someone is booked after you on the same day! In case someone is
booked on the same day, just turn the intensity turnknob of the
Argon-laser to the minimum position, switch off the key switch of the
559 nm laser, if there is at least 1 hour to next booking and leave
everything else on.
If nobody else is booked on the same day:
• Turn the Argon laser key (d) to the "Off"-position.
• Exit the control software.
• Switch off the HBO lamp.
• turn the key switch of the 559 nm laser to off and switch off the main
power.
• Switch off the laser combiner.
• Switch off the power manager (black main switch).
• Shutdown the PC.
• Switch the Argon-laser main switch off (c).