EasyManua.ls Logo

Olympus IPLEX IV9420RX - Re-Registering the Stereo Optical Adapter

Olympus IPLEX IV9420RX
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
6 Stereo measurement procedures (RX (stereo measurement function) only)
Capturing a white image
The white image affects measurement accuracy so it is necessary to adjust the brightness to the appropri-
ate level as shown below.
If dirt is on the white indented part of the check tool, it may not be able to capture an adequate white im-
age. In this case, remove the guide, which is the part with the holes for the forward-view and side-view
guides for optical adapter, from the white indented part and clean out the dirt.
The guide can be removed by turning it counterclockwise. Use a gauze cloth with a commercially avail-
able absolute ethanol or isopropyl alcohol to wipe off the dirt. To reattach the guide, repeat these steps in
reverse order.
Insufficient illumination
Optimum Excessive illumination
NOTE
7 Press the [FRZ/REC] button to freeze the white image.
Hold the insertion tube by a position behind the angulation section so that no force is applied to
the optical adapter and angulation section.
8 The following message is displayed. Remove the optical data media (USB flash
drive) from the USB connector and then select "OK" and press the [MEAS/
ENTER] joystick.
"Remove the optical data media."
Keep the optical data media (USB flash drive) in the optical adapter case so it does not get lost.
9 Check measurement values.
After the stereo optical adapter is registered, the measurement values must be checked.
See "6.5 Checking measurement values" (P. 58) for details about operations.
Applying excessive force to the optical adapter when capturing a white image may reduce measurement
accuracy.
CAUTION
After the registration of the stereo optical adapter is completed, a file with the name "IV7CALIB\FREEZE.
JPG" is created on the optical data media (USB flash drive). Removing this file does not affect operation.
NOTE
Re-register the stereo optical adapter in the same procedures as those of "Registering the new
stereo optical adapter" (P. 54).
Re-registering the stereo optical adapter
Protrusion
Bright enough that the
protrusion cannot be seen
Too bright

Table of Contents

Related product manuals