26
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-3
1. Remove the dust cap and plug from both ends of the cylinder hose for UHI-3
(MAJ-1080 only).
2. Inspect the cylinder hose for UHI-3 for damage, cracks and other
irregularities. Use the supplied wrench to attach the cylinder hose to the
CO
2
gas inlet on the rear panel of the UHI-3. Tighten with a force of
24.5 Nm (2.5 kgfm) (see Figure 3.2).
Figure 3.2
3. When using the cylinder hose for UHI-3 (MAJ-1080), attach the yoke to a
gas cylinder filled with CO
2
gas. Attach the yoke to the cylinder valve by
placing the pin of the yoke into the guide hole of the cylinder, and tighten the
handle firmly (see Figure 3.3).
Figure 3.3
24.5 Nm (2.5 kgfm)
Gas cylinder
MAJ-1080 (PIN)