Working ElementW7.074.651
5 Reprocessing of WA22366A,
WA22367A, WA22351A,
WA22355A, WA22538A
WARNING!
Infection control risk
Properly reprocess the product before rst and each
subsequent use following the instructions in this manual
and in the Olympus Endoscopy System Guide. Improper
and/or incomplete reprocessing can cause infection of the
patient and/or medical personnel. Observe the following
instrument-specic instructions.
Information in this section
The following reprocessing information supplements that
in the Olympus Endoscopy System Guide. However, if the
information given in this section differs from that in the
Olympus Endoscopy System Guide, use the information
given here.
Compatible reprocessing methods (according to
ISO 17664)
The microbiological efcacy and/or material compatibility
of the reprocessing methods listed in the table have been
validated or veried with this product. For details of the
reprocessing procedures, refer to the Olympus Endoscopy
System Guide.
++ Method validated for microbiological efcacy and
veried for material compatibility
+ Method veried for material compatibility
– Incompatible
o Contact your Olympus representative for additional
information.