36
Working Element W7.074.651
4.3 After Use
Disassemble the instrument
Remove the sheath from the working element (refer to the
sheath’s instruction manual).
Disconnect the HF cable.
Press the working element’s electrode release button.
Remove the electrode from the working element.
Press the working element’s telescope release button.
Remove the telescope from the working element.
Preparation for reprocessing at the point of use
When using the products listed below, make sure to follow
these instructions:
Immediately bring the product to the reprocessing area.
Reprocess the product following the instructions in this
manual and in the Olympus Endoscopy System Guide.
WA22366A, WA22367A, WA22351A, WA22355A,
WA22538A
Disposal
When using the products listed below, make sure to follow
these instructions*:
Dispose of the product after use.
Do not reuse or attempt to resterilize.
*Applies to the following products:
WA22301D, WA22302D, WA22305D, WA22306D,
WA22331D, WA22332D, WA22351C, WA22355C,
WA22503D, WA22507D, WA22521C, WA22523C,
WA22537D, WA22538C, WA22539D, WA22557C,
WA22558C, WA22602D, WA22603D, WA22606D,
WA22607D, WA22621C, WA22623C, WA22632D,
WA22651C, WA22655C, WA22657C