39
Working ElementW7.074.651
NOTE!
Some reprocessing methods may lead to increased wear
of the products. Inspect the products thoroughly before use
for traces of wear.
CAUTION!
Risk of damaging the product.
Do not use incompatible reprocessing methods.
Product damage may occur.
5.1 Cleaning
Immerse instruments in cleaning agent
- Do not exceed maximum recommended levels of
concentration and immersion time given by the solution’s
manufacturer.
- Make sure that no air bubbles stick to the instrument.
- All components of the instrument must be immersed
completely in the cleaning agent.
- All lumens of the instrument must be lled with cleaning
agent completely and without bubbles.
CAUTION!
Risk of damaging the instrument (working element)
Do not use any sharp, pointed objects to clean the area
indicated by the arrow.
Otherwise, the product may be damaged.