5-1
5 Troubleshooting
5.1 Hardware Issues
5.2 Error Messages
Troubleshooting
5
Symptom Possible Solution
The device does not
power up when the
power switch is on.
Make sure the battery or AC adapter is properly
connected. See section 2.11.
Erratic input signal. Check signal lines and terminal connections. Connections
must be free of corrosion. Signal lines should be
undamaged and free of sharp bends and twists. Signal
paths should avoid potential sources of noise (high
voltage and electromagnetic interference). You can also
connect the Shield ground to Earth ground to minimize
sensor noise pickup.
Symptom Possible Solution
Firmware update fails. Verify that the firmware update file is a
later release than the currently installed
firmware.