7-2
7.2 Inputs
Number of Analog Inputs: 8 or 16
Sampling Rate
Maximum Sampling Rate per Number of Channels
(Analog and Digital):
1 Channel: 125 s/sec
2 Channels: 50 s/sec
4 Channels: 25 s/sec
8 Channels: 10 s/sec
16 Channels: 5 s/sec
Logging Rate
Maximum logging Rate per Number of Channels
(Analog and Digital):
1 Channel: 125 s/sec
2 Channels: 50 s/sec
4 Channels: 25 s/sec
8 Channels: 10 s/sec
16 Channels: 5 s/sec
Specifications
7
Thermocouple
Type Range (°C)
Range (°F)
Accuracy
J -200 to 1100°C
-328 to 2012°F
±(0.15% of reading +1.1°C)
K -200 to 1370°C
-328 to 2300°F
T -200 to 400°C
-328 to 752°F
±(0.15% of reading + 1.7°C)E -200 to 1000°C
-328 to 1832°F
N -100 to 1300°C
-148 to 2372°F
R 0 to 1760°C
32 to 3200°F
±(0.15% of reading + 2.0°C)
S 0 to 1760°C
32 to 3200°F
B 500 to 1820°C
932 to 3308°F
C 0 to 2315°C
32 to 4200°F
U
W
U
U
W
U
U
W
U
U
W
U
U
W
U
U
W
U
U
W
U
U
W
U
U
W
U
Type Range (°C) Range (°F) Accuracy
Pt 100, Pt 500, Pt 1000 (0.00385 curve) -200 to 850°C -328 to 1562°F ±(0.25% of reading +1°C)
Pt 100, Pt 500, Pt 1000 (0.00392 curve) -200 to 660°C -328 to 1220°F ±(0.25% of reading +1°C)
RTD
Type Range Accuracy
2252 Ω -30 to 150°C ±1°C
10,000 Ω -5 to 150°C ±1°C
Thermistor
Range Measurement Range Accuracy
50mV -50 to 50mV
±0.1%
of FS
100mV -100 to 100mV
1V -1.00 to 1.00V
5V -5.00 to 5.00V
10V -10.00 to 10.00V
20V -20.00 to 20.00V
50V -50.00 to 50.00V
Voltage
Range Measurement Range
20mA -20 to 20mA ±0.1% of FS
Current
Measurement Range Accuracy
0 to 250 KHz ±2 Hz
Frequency