31
Guidance and manufacturer’s declaration - electromagnetic emissions
OMRON BP7100 is intended for use in the electromagnetic environment specified below. The
customer or the user of this OMRON BP7100 should assure that it is used in such environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The OMRON BP7100 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The OMRON BP7100 is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not Applicable.
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Not Applicable.
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP7100 is intended for use in the electromagnetic environment specified below. The
customer or the user of this OMRON BP7100 should assure that it is used in such environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Not Applicable. Not Applicable.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Not Applicable. Not Applicable.